Emerson X-Stream gas analyzer

The X-STREAM™ series of gas analyzers features multi-channel analysis utilizing infrared, ultraviolet and visible (NDIR/UV/VIS) photometry, paramagnetic and electrochemical oxygen, and thermal conductivity sensor technologies.

Please read the PDF-file(s) above for more detailed information.

X-STREAM analyzers can measure up to 4 components and the measuring principles may be ordered in various combinations. The physical benches can be installed in their own compartment separated from the electronics. A purge is added for handling corrosive and toxic gases to protect the electronics and to provide operator safety. Optional thermostatic control enables measuring lower sample gas concentrations and measurements at higher dew points.

X-STREAM analyzers, equipped with an internal wide range power supply for all world areas, offer analog outputs, 4 status signal relay outputs (according to NAMUR NE 107) and Modbus communication over an Ethernet or serial interface.

The analyzers can be ordered with either screw terminals or connectors for electrical connections. Supplied as tabletop version, the analyzer can be modified to a rack mount version by an accessory kit.

The instrument has an alphanumeric display and is operated manually by six keys without the need to open the enclosure. Clear text messages (available in several languages) and front panel LED provide information about the measurement and analyzer status.

Features
X-STREAM (Standard Model)

  • Analyzes up to four components
  • New IntrinzXTM photometric technology with intrinsic linearity provides high sensitivity, large dynamic ranges and long-term span stability with extended calibration intervals
  • User-friendly operator interface
  • Easy to install, low maintenance and field repairable
  • Supports NDIR, UV, VIS, paramagnetic and electrochemical O2, and thermal conductivity detectors
  • Solvent-resistant, corrosion-resistant, intrinsically safe and “infallible containment” solutions are available
  • Analog and relay outputs, digital inputs, Ethernet and serial interface with Modbus communication
  • Autocalibration via internal or external valves
  • Optional barometric pressure compensation, internal sampling pump and flow sensor

X-STREAM XE (Enhanced Model)

  • Analog inputs
  • Web browser access
  • Data, calibration, and event logger
  • Support for e-mail alarm & event messaging, and USB storage devices
  • Calculator for virtual measurements
  • PLC for control of sample handling and sample lines